Single Pass Aseptic Isolators

isolator_api

Traditional aseptic isolators suffer from several design shortcomings. Double-skinned rear walls make them expensive to manufacture. They're also difficult to decontaminate and require long aeration times after decontamination. Plus, installing safe-change outlet filters can be difficult.

MBRAUN has addressed these shortcomings in a single-pass aseptic isolator system. Single-pass design eliminates the need for double-skinned walls for return airflow.

The working area of these isolators allows direct connection of inlet and outlet filters, minimizing contamination and simplifying validation. Aeration times are quicker because of the units' low volume and 100% air exchange. Safe-change filters are standard.

MBRAUN single-pass aseptic isolator features:

  • Operate at +200 pascals
  • Unidirectional flow of 0.20 to 0.45 meter per second
  • Uniform down-flow via compressed gas diffuser
  • Tolerant of hydrogen-peroxide steam decontamination
  • Hinged window

Please contact us if you have questions choosing the right single pass aseptic isolator system for your specification needs.

Features

  • The system is operated at a relative positive pressure of +200 Pa +/ pressure.
  • The isolator is equipped with a filter system to protect the outside from materials used inside the isolator.
  • Can be equipped with a bio decontamination (sterilization) system.
  • The main chamber and the transfer device(s) can be decontaminated (sterilized) separately or parallel.
  • The Isolator is equipped with a Safety and Control on the Isolator.
  • The transfer device is equipped with 2 doors.
  • Pressure indicated by a Magnehelic gauge with calibration certificate
  • Pressure measurement by pressure sensor with trend indication on the user panel
  • Trending capabilities of system parameters.
  • Manual leak test.
  • Differential pressure monitoring of the filters.
  • Air stream sensor (air exchange rate)

Options

Please contact MBRAUN for more information.

Specifications

Outside dimensions (including the transfer device, covers and stand):
Width 2570 mm; Height 2600 mm; Depth 900 mm

Isolator Body:
Stainless steel 1.4404 (316L)
Isolator Stand: Stainless steel 1.4404 (316L)
Window Frame: Stainless steel 1.4404 (316L)
Window: Laminated Safety Glass
Glove Ports: HDPE with FDA certificate
Gloves: HDPE with FDA certificate
Inlet Filter: HEPA H14 filter with certificate (Filter change from the outside)
Outlet Filter: Safe change HEPA H14 filter with certificate (filter change from inside)
___________________________________________________________________________

Dimensions of the Transfer Device
Width 400 mm; Height 450 mm; Depth 490 mm

Transfer Device Body:
Stainless steel 1.4404 (316L)
Transfer Device Door: Safety glass and stainless steel frame 1.4404 (316L)